commands:ensync

Command:ENSYNC (ES)

MS2000 or RM2000 syntax
Shortcut ES
Format ENSYNC [axis]=[position in mm]…
Units position in millimeter
Remembered Using SS Z
Tiger syntax
Shortcut ES
Format ENSYNC [axis]=[position in mm]…
Units position in millimeter
Type Axis-Specific
Remembered Using [addr#]SS Z

This command lets the user set a position, in millimeters - absolute, which will toggle a TTL output when the stage crosses that position. When ENSYNC is issued, the TTL output is reset low. Whenever the stage crosses the ENSYNC position, the output will toggle low to high and if crossed again, from high to low. ENSYNC will only work with one axis at a time, either X or Y and depends on how JP1 is jumped.

On MS2000

(JP1-1&2 = X axis, JP1 – 2&3 = Y axis) The TTL output is available on pin SV1-7. Contact ASI for additional details on these modifications

On Tiger

Need Dual Axis card Rev F3 and above. Pin is routed to the backplane. Additional hardware is needed to expose the signal. Contact ASI for additional details on these modifications

Warning—units of the position info is millimeters rather than tenths of microns