Tunable Lens 4F Assembly leverages Tunable Lens to remotely focus an image on a microscope (ie without moving the objective itself).
For example on a ASI RAMM Infinity microscope, the following objectives caused the varying Z focus change.
^Objective Magnification ^Z focus change for full 4000 units travel ^
| 5x | 1600um |
| 20x | 80-90um |
| 40x | 50um |
| 50x | 15um |
| 60x | 8um |
The relation between objective magnification and focus change is
\begin{equation}Focus Change = \frac{-1}{M_{obj}^2}*\frac{n*f_r^2}{f_{ETL}}\end{equation}
where Mobj is objective magnification; n is refractive index of immersion medium; fr is focal length of the two lenses in 4F assembly; FETL is effective focal length of the ETL in combination with any offset lens.
As described in Fahrbach, Florian O., et al. “Rapid 3D Light-Sheet Microscopy with a Tunable Lens.” Optics Express, vol. 21, no. 18, 2013
Tunable Lens are susceptible to temperature change, their diopter decrease as temperature increases.Below is a graph of diopter vs current at two different temperatures
This diopter per celcius change isn't constant and varies too. Below is a graph of the change diopter change per celsius vs current.
Fortunately this effect is predictable and the manufacturer has built a temperature sensor into the Tunable lens and provided characterization data. At the factory , we analyze this data and build a model. The parameters for this model are saved on the Tunable Lens EEPROM itself. The Tunable Lens card reads the EEPROM on starup. These parameters can be read and altered thru serial commands ERROR(E), PCROS(PC), PG and PSG
When Temperature compensation is enabled (thru the PM [Axis]=2) command, firmware periodically reads the temperature sensor on the Tunable Lens and adjust the DAC. This change won't be reflected in the position of the Tunable Lens read with WHERE(W) or set with MOVE (M). The correction is happening on a internal layer , right before the DAC and current amplifier.
Below is how the temperature compensation is calculated and applied.
\begin{equation}\frac{D}{T} = I_{user}*K_1+C_1\end{equation}
\begin{equation}D = \frac{D}{T}*(T_{current}-T_{setpoint})\end{equation}
\begin{equation}I_{applied} = I_{user}- D*K_{D2I}\end{equation}
The coefficients K1 , KD2I and constants C1 and Tsetpoint are calculated from two Diopter vs Current curves the Tunable lens manufactured provides with each Tunable lens. They are calculated by ASI technicians and stored onboard the tunable lens itself. If a user would like to alter any of these settings , they may do it thru the following commands.
\begin{equation} X=\frac{K_1*-1}{10000000} \end{equation}
One of the main application of Tunable Lens system is with ASIs XYZ Tracker plugin as a continuous focus device. For more info refer ASIs XYZ Tracker
In order to add a CRISP autofocus device to the simplest form of ASI's inverted Modular Infinity Microscope1), an additional optical path must be installed on the microscope, a control card must be inserted into the Tiger controller and plugged into the card of the axis it will control (typically the Z-axis), and the firmware of that card must be updated.
1)Click Step 2 “Dither” state. Adjust the main CRISP adjusting Thumb Screw for maximum ERR number. 2)Click Step 3 “Set Gain” and wait a few seconds for it to finish. 3)Re-focus microscope on sample and set the Z-axis controller position to Zero. (HERE Z=0) 4)Be sure NA setting is correct for the objective and click “GRAPH” button.
5)Choose and operating point and adjust the main CRISP adjusting Thumbscrew to get there. If you only adjust for maximum ERR number, then the point you choose will be at the steepest part of the focus curve. Sometimes this can be very close to where the curve makes a sudden change in slope.
The MPPI-3 Milli Pulse Pressure Injector, has been designed and tested to provide years of reliable service.
Note: In order to insure that the unit operates as designed, use only nitrogen or clean dry compressed air as the pressure source.
The design of the MPPI-3 requires little maintenance on the part of the user. In fact, there are only two areas that the user needs to be aware. The first is to periodically check for moisture and other contaminants in the pressure source. The internal in-line filter that protrudes from the back of the unit has a small needle valve located on the bottom. This valve should be depressed periodically to check for moisture from the pressure source. If excessive moisture is noted, the pressure source should be changed. In the event that the filter becomes degraded or plugged, contact ASI for a replacement.
The second area of maintenance that the user should be aware is the replacement of the fuse. In the event that the MPPI-3 fails to turn on, the line fuse located inside the MPPI should be checked. If the fuse is blown, an exact replacement needs to be used. The MPPI-3 uses a 250mA fast-blow fuse. If the fuse blows again after replacement, please contact ASI for instructions on returning the unit for repair.
<html> <iframe width=“560” height=“315” src=“https://www.youtube.com/embed/o4qVIpkSC_U?start=02&end=122” frameborder=“0” allow=“autoplay; encrypted-media” allowfullscreen></iframe> </html>
end of mg section