The ASI PhotoTrack solves the problem of tracking a moving specimen. Combined with ASI’s closed loop DC servo motor XY stages, PhotoTrack allows you to keep your free-moving samples in your field of- view (FOV) no matter where they want to go. PhotoTrack uses a low-light-level rapid position sensing quadrant PMT sensor to monitor a right reference spot on the target. It then commands the XY stage to maintain that spot in the center of a microscope’s field-of-view.
| Sensor Mount | Parfocal C-Mount |
| Sensor Controls | Manual Gain, On/Off, Iris, Overload Reset |
| Beam Splitter Ratio | 20/80, 10/90, or Dichroic |
| PMT Type | Hamamatsu 4-channel XY quadrant |
| PMT Protection | High voltage shutdown on overload |
| PMT Usable Dynamic Range | > 60 dB |
| Servo Loop Time | 3 milliseconds |
| Read Time | 2 milliseconds |
| Overall Response Time | < 25 milliseconds |
| Sensor Dimensions (H x W x D, not counting plug) |
74 x 94 x 79 millimeters 2.9 x 3.7 x 3.1 inches |
| Controller for PhotoTrack Sensor | Integrated with specially-configured ASI MS-2000 controllers |
Worm-tracking challenge leads to new tool for brain research
Copyright © Applied Scientific Instrumentation 2001 - 2013. All rights reserved.