Command:PR

This is a Tiger only command, that has different usage for MicroMirrors and Piezos.

MicroMirror
Shortcut PR
Format PR [axis]=[5 to 10] …
Units Integer code, 5 to 10
Type Axis-Specific
Remembered Automatically
Firmware Requiredv2.83+

This command is “recycled” for a different use in MicroMirror axes than for piezo. In the context of a MicroMirror axis this command is used to set the MicroMirror travel range. Settings is automatically saved into non-voltatile memory, however controller needs a system RESET or RESTART for setting to take effect.

PR[Axis Name] = MicroMirror Range
in degrees
5 5
6 6
8
(default)
8
10 10

Example

PR A=5		
:A

Sets range of A axis as 5 degrees

PR A? B?			
A=5 B=8 :A

Queries the range of axes

Piezo
Shortcut PR
Format PR [axis]=[0 to 8] …
Units Integer code, 0 to 8
Type Axis-Specific
Remembered Automatically

PR is used to set the piezo travel range. It is an Axis specific command. Setting is automatically saved in the non-volatile memory. Will need a system RESET or RESTART for setting to take effect.

PR [Axis Name] = Piezo Range
in microns
1 50
2 100
3 150
4 200
5 300
6 350
7 500
8 70
Tunable Lens
Shortcut PR
Format PR [axis]=[0 to 3] …
Units Integer code, 0 to 3
Type Axis-Specific
Remembered Automatically

PR is used to set the Tunable Lens units or user input. It is an Axis specific command. Setting is automatically saved in the non-volatile memory. Will need a system RESET or RESTART for setting to take effect. For more info refer to Units/Resolution section on TGTLC card page.

PR [Axis Name] = Units
1 (default) [0 to 290mA] with coordinates -32768 to +32768
2 [0 to 290mA] 1/1000 of dpt, range usually between 26000 to 7000
3 [-290mA to +290mA] with coordinates -32768 to +32768
SIGNAL_DAC for TGDAC4 / TGGALVO
Shortcut PR
Format PR [axis]=[0-7] …
Units Integer code, 0-7
Type Axis-Specific
Remembered automatically, no SS Z required
Firmware Requiredv3.3+

For TGDAC4 card with SIGNAL_DAC firmware, the bits 0-2 of the specified code sets the output range (values 0-7 in decimal). Bits 3-7 are reserved for future use. Controller reset or restart is needed for setting to take effect.

For the SIGNAL_DAC firmware, the axis units are always in millivolts (0.001V), regardless of the PR setting. For example, when PR H=2, the maximum axis value of H is 10240. When PR H=0, then the maximum axis value of H is 2048.

Code (Decimal) Code (Binary) Output range
0 000 0V to 2.048V
1 001 0V to 4.096V
2 010 0V to 10.24V
3 011 not supported
4 100 -1.024V to 1.024V
5 101 -2.048V to 2.048V
6 110 -5.12V to 5.12V
7 111 -10.24V to 10.24V

Example

PR A=2 B=7
:A

Puts the output A in 0-10V mode and output B in +/- 10V mode.

PR A? B?			
:A A=2 B=7 