| Shortcut | AL |
|---|---|
| Format | AFLIM [X=Log_amp_AGC] [Y=LED_intesity_pot] [Z=in_focus_mm] |
| Remembered | Using SS Z |
| Shortcut | AL |
|---|---|
| Format | [Addr#]AFLIM [X=Log_amp_AGC] [Y=LED_intesity_pot] [Z=in_focus_mm] |
| Type | Card-Addressed |
| Remembered | Using [Addr#]SS Z |
X/Y: The X and Y arguments of this command to directly read and write values (0 to 255) to the CRISP electronics digital potentiometers. Not recommended for use with host software.
AL X.AL Y.
Z [in_focus_mm]: The Z-argument specifies the focus precision (in millimeters) when the lock state changes from K or k to F. Useful for automatic checking of desired focus stability. Also useful to enforce a tighter or looser focus state before indicating a lock condition. Note that this value is overwritten whenever the NA of the objective is specified via the LR Y command as of November 2015.
| Shortcut | AL |
|---|---|
| Format | AFLIM [X= x-axis highlight] [Y= y-axis highlight] [Z= safety limit enable] |
| Remembered | Using SS Z |
| Shortcut | AL |
|---|---|
| Format | [Addr#]AFLIM [X= x-axis highlight] [Y= y-axis highlight] [Z= safety limit enable] |
| Type | Card-Addressed |
| Remembered | Using [Addr#]SS Z |
X/Y: The X and Y values set the length and breadth of the Sampled/Highlighted Video area. Range is 0 to 100, with the value of 0 covering 0% of the video frame and 100 covering 90% of video frame.
Z: The Z value enables or disables the 200 µm safety limit described in the AUTOFOCUS OPERATION section on page 4. Setting safety limit enable = 1 enables the safety limit; safety limit enable = 0 disables the safety feature. The default value is 1.
Warning! Disabling the safety limit could result in damage to your optics, your sample, or your focus drive.
AL X=80 Y=50 Z=1 :A<CR><LF>
AL :N-3
Error indicates missing arguments
AL X=1000 Y=-12 :N-4
Error indicates arguments out of range
AL X=90 Y=90 :N-5
Error indicates operation failed, try entering one argument at a time
AL X? Y? Z? :A X=80 Y=50 Z=1