====== Command:SETUP (SU) ====== --> MS2000 or RM2000 syntax ^ ^Shortcut |SU| ^Format |SU [axis]=[position in mm]…| ^Units |Millimeters| ^Remembered |Automatically | <-- --> Tiger syntax ^ ^Shortcut |SU| ^Format |SU [axis]=[position in mm]…| ^Units |Millimeters| ^Type |Axis-Specific | ^Remembered |Automatically | <-- Same as [[commands:setlow|SETLOW command]] (see above) but for upper firmware limit switch. **Note 1:** If this value is equal to or less than the value for ''SETLOW'', then the controller will operate incorrectly. See also Note 2.\\ **Note 2:** When the direction of an axis is negative (see [[commands:customa|CCA Z=###]]), upper limit settings must be negative values, and lower limit settings must be positive values. \\ **Note 3:** For clocked devices (e.g. filter sliders, turrets) the upper limit is always the number of positions. Querying the value is useful for determining how many positions. \\ **Note 4:** As of Tiger firmware v2.8, ''SU [axis]+'' will set the upper limit to be the current position. Restore the default limit by executing'' SU [axis]-''. \\ {{tag>commands tiger ms2000}}