====== Command:SAVESET (SS) ====== --> MS2000 or RM2000 syntax^ ^Shortcut |SS| ^Format |SAVESET [X][Y][Z] | <-- --> Tiger syntax^ ^Shortcut |SS| ^Format |[addr#]SAVESET [X][Y][Z] | ^Type |Card-Addressed | <-- ''SAVESET'' allows the user to save current parameters settings to Flash memory. \\ ''SAVESET Z'', saves settings to flash memory \\ ''SAVESET Y'', restores previously saved settings after a ''SAVESET X'' \\ ''SAVESET X'', will reload factory defaults upon next power-up \\ ==Reply== Upon the start of execution of this command, the controller will reply with a : . When the execution is complete, an A will follow the colon. **Note 1:** During the time interval between the : and the A , no serial or manual moves should be given. **Note 2:** In MS2000 Versions 6.1u and later (see ''VERSION'' command), limit settings (see ''SETLOW'' and ''SETUP'') are saved if and only if the ''SAVEPOS'' command is issued after the command ''SAVESET Z''. == Example == --> MS2000 example ^ $ SS Z :A Save the current settings to flash memory. <-- --> Tiger example ^ $ 1SS Z :A Save the Tiger Card at Card Address 1's current settings to flash memory. <-- {{tag>commands tiger ms2000}}