====== Command:SAVEPOS (SP) ====== --> MS2000 or RM2000 syntax ^ ^Shortcut |SP | ^Format |SP [X=inhibit]| ^Units |0 or 1 only| ^Remembered |Using SS Z | <-- --> Tiger syntax ^ ^Shortcut |SP | ^Format |[addr#]SP [X=inhibit]| ^Units |0 or 1 only| ^Type |Card-Addressed| ^Remembered |Using [addr#]SS Z | <-- The axis positions and soft limit locations are usually automatically saved when power is turned off. If this action is not desired, setting inhibit=1 will prevent power down saves. (Default is ''inhibit = 0'') If the command is given without argument, a save position shutdown will be initiated whereby the axes will be halted, positions saved to flash memory, and the controller placed in a non-responsive condition until power is cycled. == Reply == Upon the start of execution of this command, the controller will reply with a : . When the execution is complete, an A will follow the colon. When a power down condition is detected, an O is transmitted. After the positions are successfully saved, a K is sent. **Note 1:** During the time interval between the : and the A , no serial or manual moves should be given. **Note 2:** In MS2000 Versions 6.1u and later (see ''VERSION'' command), limit settings (see'' SETLOW'' and ''SETUP'') are saved if and only if the ''SAVEPOS'' command is issued after the command ''SAVESET Z''. **Note 3:** OK is not transmitted in TG-1000 {{tag>commands tiger ms2000}}