====== Command:AFLIM (AL) ======
--> For CRISP ^
--> MS2000 and RM2000 Syntax ^
^Shortcut |AL|
^Format |AFLIM [X=Log_amp_AGC] [Y=LED_intesity_pot] [Z=in_focus_mm] |
^Remembered |Using SS Z |
<--
--> Tiger Syntax ^
^Shortcut |AL|
^Format |[Addr#]AFLIM [X=Log_amp_AGC] [Y=LED_intesity_pot] [Z=in_focus_mm] |
^Type |Card-Addressed|
^Remembered |Using [Addr#]SS Z |
<--
**X/Y:** The X and Y arguments of this command to directly read and write values (0 to 255) to the CRISP electronics digital potentiometers. Not recommended for use with host software.
* See the [[:commands:lock|LK M]] command to adjust the LogAmp_AGC instead of ''AL X''.
* See the [[:commands:unlock|UL X]] command to adjust the LED intensity instead of ''AL Y''.
\\
**Z** ''[in_focus_mm]'': The Z-argument specifies the focus precision (in millimeters) when the lock state changes from K or k to F. Useful for automatic checking of desired focus stability. Also useful to enforce a tighter or looser focus state before indicating a lock condition. Note that this value is overwritten whenever the NA of the objective is specified via the LR Y command as of November 2015.
<--
--> For Video Autofocus ^
--> MS2000 and RM2000 Syntax ^
^Shortcut |AL|
^Format |AFLIM [X= x-axis highlight] [Y= y-axis highlight] [Z= safety limit enable] |
^Remembered |Using SS Z |
<--
--> Tiger Syntax ^
^Shortcut |AL|
^Format |[Addr#]AFLIM [X= x-axis highlight] [Y= y-axis highlight] [Z= safety limit enable] |
^Type |Card-Addressed|
^Remembered |Using [Addr#]SS Z |
<--
**X/Y:** The X and Y values set the length and breadth of the Sampled/Highlighted Video area. Range is 0 to 100, with the value of 0 covering 0% of the video frame and 100 covering 90% of video frame.
**Z:** The Z value enables or disables the 200 µm safety limit described in the AUTOFOCUS OPERATION section on page 4. Setting safety limit enable = 1 enables the safety limit; safety limit enable = 0 disables the safety feature. The default value is 1.
**Warning!** Disabling the safety limit could result in damage to your optics, your sample, or your focus drive.
$ AL X=80 Y=50 Z=1
:A
$ AL
:N-3
Error indicates missing arguments
$ AL X=1000 Y=-12
:N-4
Error indicates arguments out of range
$ AL X=90 Y=90
:N-5
Error indicates operation failed, try entering one argument at a time
$ AL X? Y? Z?
:A X=80 Y=50 Z=1
<--
{{tag>commands tiger ms2000 crisp autofocus}}